ART19306 Rev. A 12/09
Operators Manual
Blood Glucose and β-Ketone Monitoring System
Precision Xceed Pro, TrueID, TrueMeasure, TrueAccess, PrecisionWeb, Precision, MediSense,
Optium and Precision PCx are trademarks of the Abbott Group of Companies in various
jurisdictions. Other trademarks are the property of their respective owners.
This product(s) and/or its use are protected by one or more of the following patents:
US5,509,410; US5,628,890; US5,727,548; US6,129,823; US6,592,745; US6,607,658;
US6,736,957; US6,764,581; US6,939,450; US6,377,894; US6,600,997; US6,773,671;
US5,682,884; USD7,077,328; USD411,989; USD413,537; USD557,422S; US6,591,125;
US7,058,437; EP1,009,850B1; EP1,119,637B8; EP1,023,455B1; and EP1,012,326B1.
Additional patents may be issued and/or pending.
© 2009 Abbott
Abbott Diabetes Care Inc.
1360 South Loop Road
Alameda, CA 94502 USA
Vista de pagina 0
1 2 3 4 5 6 ... 84 85

Indice de contenidos

Pagina 1 - Operator’s Manual

ART19306 Rev. A 12/09Operator’s Manual Blood Glucose and β-Ketone Monitoring SystemPrecision Xceed Pro, TrueID, TrueMeasure, TrueAccess, PrecisionWeb,

Pagina 2 - Table of Contents

2. Using the Monitor Bar Code Scanner Data Entry Keypad Audible Indicator Data Port Battery Compartment Port Protector Monitor Identification

Pagina 3

2-1Questions? Refer to Page 10-10 for Technical Support Instructions.Using the MonitorBar Code ScannerThe bar code scanner enables you to scan the inf

Pagina 4 - 1. Overview

2-2Questions? Refer to Page 10-10 for Technical Support Instructions.On/Off Press this button to turn the monitor on. Each time the monitor is turned

Pagina 5 - Overview

2-3Questions? Refer to Page 10-10 for Technical Support Instructions.Monitor IdentificationEach monitor has a unique serial number listed on the back

Pagina 6 - Conventions

3. Patient Test Operating Guidelines for All Samples Collecting Blood Samples How to Obtain a Capillary Blood Sample How to Obtain a Venous Bl

Pagina 7 - System Components

3-1Questions? Refer to Page 10-10 for Technical Support Instructions.This chapter describes the steps to be taken to test patient blood glucose or β-k

Pagina 8 - Blood β-Ketone Test Strip

3-2Questions? Refer to Page 10-10 for Technical Support Instructions.Collecting Blood Samples How to Obtain a Capillary Blood Sample:Use only fresh wh

Pagina 9 - Chapter 1: Overview

3-3Questions? Refer to Page 10-10 for Technical Support Instructions.Patient Test Procedure - GlucoseUse the following procedure to perform a patient

Pagina 10 - 2. Using the Monitor

What You See on the Display What You Do Comments3-4Questions? Refer to Page 10-10 for Technical Support Instructions.Press 2 to Conrm Patient ID or 1

Pagina 11 - Chapter 2: Using the Monitor

What You See on the Display What You Do Comments3-5Questions? Refer to Page 10-10 for Technical Support Instructions.12. If required, scan or manuall

Pagina 12 - Port Protector

iQuestions? Refer to Page 10-10 for Technical Support Instructions.This Operator's Manual provides basic information about the Precision Xceed Pr

Pagina 13 - Menu Tree

3-6Questions? Refer to Page 10-10 for Technical Support Instructions.What You See on the Display What You Do Comments13. You can select one of the fo

Pagina 14 - 3. Patient Test

3-7Questions? Refer to Page 10-10 for Technical Support Instructions.Patient Test Procedure - β-KetoneUse the following procedure to perform patient t

Pagina 15 - Patient Test

4. Control Test Operating Guidelines for Control Solution Testing Control Test Procedure - Glucose Control Test Procedure - β-Ketone

Pagina 16 - Disposing of Waste

4-1Questions? Refer to Page 10-10 for Technical Support Instructions.This chapter describes the steps for running a control test with control solution

Pagina 17 - Chapter 3: Patient Test

4-2Questions? Refer to Page 10-10 for Technical Support Instructions.Note: The following section illustrates some common settings. Depending on your i

Pagina 18

What You See on the Display What You Do Comments4-3Questions? Refer to Page 10-10 for Technical Support Instructions.5. P r e s s Scan to scan the

Pagina 19

What You See on the Display What You Do Comments4-4Questions? Refer to Page 10-10 for Technical Support Instructions.9. Wait for the monitor to analy

Pagina 20

4-5Questions? Refer to Page 10-10 for Technical Support Instructions.What You See on the Display What You Do Comments1. Repeat steps 1 through 6 from

Pagina 21

5. Data Review Data Review for Patient by Operator ID Data Review for Patient by Patient ID Data Review for All Patient Data Data Review for Contr

Pagina 22 - 4. Control Test

5-1Questions? Refer to Page 10-10 for Technical Support Instructions.The Precision Xceed Pro Monitor stores the results of up to 2,500 patient tests,

Pagina 23 - Control Test

iiQuestions? Refer to Page 10-10 for Technical Support Instructions.10. Troubleshooting ...

Pagina 24 - Chapter 4 : Control Test

5-2Questions? Refer to Page 10-10 for Technical Support Instructions.4. P r e s s Scan to scan the Operator ID bar code or manually enter the Ope

Pagina 25

5-3Questions? Refer to Page 10-10 for Technical Support Instructions.Data Review for Patient by Operator IDPlease refer to the beginning of this Data

Pagina 26

5-4Questions? Refer to Page 10-10 for Technical Support Instructions.105 mg/dL1. P r e s s 2 to select Patient by PatID. 2. Press Scan to scan th

Pagina 27

5-5Questions? Refer to Page 10-10 for Technical Support Instructions.1. P r e s s 3 to select All Patient Data. The monitor will show all patient

Pagina 28 - 5. Data Review

5-6Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 4 to select Control Data.The monitor shows the result of the most recen

Pagina 29 - Data Review

5-7Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 5 to select Proficiency Data. The monitor shows the most recent profici

Pagina 30 - Chapter 5: Data Review

5-8Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 6 to select Linearity Data. The monitor shows linearity tests first by

Pagina 31 - 105 mg/dL

6. Proficiency TestProficiency Test - GlucoseProficiency Test - β-Ketone

Pagina 32

6-1Questions? Refer to Page 10-10 for Technical Support Instructions.Proficiency Test - GlucosePlease use the following procedures to perform a profic

Pagina 33

What You See on the Display What You Do Comments6-2Questions? Refer to Page 10-10 for Technical Support Instructions.6. P r e s s Scan to scan the

Pagina 34 - Data Review for Control Data

1. Overview Intended Use TrueID, TrueMeasure and TrueAccess Icons and Warnings Conventions System Components

Pagina 35

What You See on the Display What You Do Comments6-3Questions? Refer to Page 10-10 for Technical Support Instructions.9. Bring prociency survey speci

Pagina 36 - 45 mg/dL

6-4Questions? Refer to Page 10-10 for Technical Support Instructions.1. Follow steps 1 through 8 in the last section on the Glucose Proficiency Test.

Pagina 38 - Proficiency Test

7-1Questions? Refer to Page 10-10 for Technical Support Instructions.What You See on the Display What You Do CommentsLinearity Test1. P r e s s On/

Pagina 39 - Chapter 6: Proficiency Test

What You See on the Display What You Do Comments7-2Questions? Refer to Page 10-10 for Technical Support Instructions.5. Select Glucose or β-Ketone Li

Pagina 40

What You See on the Display What You Do Comments7-3Questions? Refer to Page 10-10 for Technical Support Instructions.9. P r e s s Scan to scan the

Pagina 41 - Proficiency Test - β-Ketone

What You See on the Display What You Do Comments7-4Questions? Refer to Page 10-10 for Technical Support Instructions.12. Follow the instructions in t

Pagina 42 - 7. Linearity Test

8. Review Setup Monitor Configuration Review Setup Review Test Setup for All Tests Review Test Setup for Test Type Review Test Setup for Patient

Pagina 43 - Linearity Test

8-1Questions? Refer to Page 10-10 for Technical Support Instructions.Monitor ConfigurationThe Precision Xceed Pro Monitor configuration settings are e

Pagina 44 - Chapter 7: Linearity Test

8-2Questions? Refer to Page 10-10 for Technical Support Instructions.Operator IDDecide whether an operator can continue to test (“allow test” or “warn

Pagina 45

1-1Questions? Refer to Page 10-10 for Technical Support Instructions.Intended UseThe Precision Xceed Pro Point of Care System for Blood Glucose and β-

Pagina 46

8-3Questions? Refer to Page 10-10 for Technical Support Instructions.1. P r e s s On/Off to turn on the monitor.The Abbott logo screen will appear

Pagina 47 - 8. Review Setup

8-4Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 1 for Test Setup.2. P r e s s 1 for All Tests to continue to the All

Pagina 48 - Review Setup

8-5Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 1 for Test Setup.2. P r e s s 2 for Test Type to continue to the Test

Pagina 49 - Chapter 8: Review Setup

8-6Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 1 for Test Setup.2. Press 3 for Patient Test to continue to the Patient

Pagina 50

8-7Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 1 for Test Setup.2. Press 4 for Control Test to continue to the Control

Pagina 51

What You See on the Display What You Do Comments8-8Questions? Refer to Page 10-10 for Technical Support Instructions.7. P r e s s 1-Next to contin

Pagina 52

8-9Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press 1 for Test Setup.2. P r e s s 5 for Security to continue to the Secur

Pagina 53

What You See on the Display What You Do Comments8-10Questions? Refer to Page 10-10 for Technical Support Instructions.7. P r e s s 1-Next to contin

Pagina 54

8-11Questions? Refer to Page 10-10 for Technical Support Instructions.1. Press On/Off to turn on the monitor.The Abbott logo screen will appear in bl

Pagina 55 - Test Setup menu

8-12Questions? Refer to Page 10-10 for Technical Support Instructions.6. P r e s s 1-Next to continue to the System Status Time screen.7. P r e

Pagina 56

1-2Questions? Refer to Page 10-10 for Technical Support Instructions. Chapter 1: OverviewIcons and WarningsInformation that is of particular relevance

Pagina 57

9. Data Upload Docking Station Mounting Instructions Enabling Data Upload

Pagina 58 - System Status

9-1Questions? Refer to Page 10-10 for Technical Support Instructions.Once the test results have been collected during a prescribed period of time, the

Pagina 59

9-2Questions? Refer to Page 10-10 for Technical Support Instructions.Docking Station The Precision Xceed Pro System offers an optional docking station

Pagina 60 - 9. Data Upload

9-3Questions? Refer to Page 10-10 for Technical Support Instructions.Enabling Data UploadThe docking station has two ports located on the back: an AC

Pagina 61 - Data Upload

10. Troubleshooting Troubleshooting Patient Test Results Troubleshooting Out-of-Range Control Test Results Troubleshooting Out-of-Range Linearity

Pagina 62 - Mounting Instructions

10-1Questions? Refer to Page 10-10 for Technical Support Instructions.Troubleshooting Patient Test ResultsThis section describes conditions that can c

Pagina 63 - Enabling Data Upload

10-2Questions? Refer to Page 10-10 for Technical Support Instructions.When You Turn on the Monitor, or During Use:The display is blank.Symptom/Error M

Pagina 64 - 10. Troubleshooting

10-3Questions? Refer to Page 10-10 for Technical Support Instructions.When You Turn on the Monitor, or During Use:Symptom/Error Message Explanation Ho

Pagina 65 - Troubleshooting

10-4Questions? Refer to Page 10-10 for Technical Support Instructions.When You Turn on the Monitor, or During Use:Symptom/Error Message Explanation Ho

Pagina 66 - Error Messages

Symptom/Error Message Explanation How To Respond10-5Questions? Refer to Page 10-10 for Technical Support Instructions.The Glucose QC Interval Controls

Pagina 67 - Chapter 10: Troubleshooting

1-3Questions? Refer to Page 10-10 for Technical Support Instructions.System Components The Precision Xceed Pro Monitor is part of an overall system de

Pagina 68

10-6Questions? Refer to Page 10-10 for Technical Support Instructions.Symptom/Error Message Explanation How To RespondDuring Any Test:The bar code jus

Pagina 69 - When a Test is Selected:

10-7Questions? Refer to Page 10-10 for Technical Support Instructions.During Any Test:The Operator ID just entered is not on the list of acceptable O

Pagina 70 - During Any Test:

10-8Questions? Refer to Page 10-10 for Technical Support Instructions.Symptom/Error Message Explanation How To RespondThe test result is below the mea

Pagina 71

10-9Questions? Refer to Page 10-10 for Technical Support Instructions.When a Control Test is Selected:When a Linearity Test is Selected:A control solu

Pagina 72 - 20 mg/dL

10-10Questions? Refer to Page 10-10 for Technical Support Instructions.Technical Support InstructionsAbbott Technical Support contact information is l

Pagina 73 - When Data Review is Selected:

10-11Questions? Refer to Page 10-10 for Technical Support Instructions.Contacting Abbott for ServiceAbbott is committed to helping you resolve any pro

Pagina 74 - Returning an Instrument

10-12Questions? Refer to Page 10-10 for Technical Support Instructions.LATIN AMERICAArgentinaAbbott Laboratorios de Argentina S/AAbbott Diabetes CareB

Pagina 75 - Contacting Abbott for Service

11. Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries

Pagina 76 - NORTH AMERICA

11-1Questions? Refer to Page 10-10 for Technical Support Instructions.IMPORTANT: Do not allow liquid to enter the monitor. Permanent damage could occu

Pagina 77 - 11. Maintenance

11-2Questions? Refer to Page 10-10 for Technical Support Instructions.Battery Life The average alkaline battery life is approximately 60 days, based o

Pagina 78 - Maintenance

1-4Questions? Refer to Page 10-10 for Technical Support Instructions. Chapter 1: OverviewTest StripsPrecision Xceed Pro Test Strips offer the latest a

Pagina 79 - Replacing the Batteries

12. SpecificationsPrecision Xceed Pro MonitorSymbolsBar Code TypesImportant Safety Instructions

Pagina 80 - 12. Specifications

12-1Questions? Refer to Page 10-10 for Technical Support Instructions.Precision Xceed Pro MonitorDimensions:Length: 19.7 cm (7.7 in)Width: 7.5 cm (

Pagina 81 - Specifications

12-2Questions? Refer to Page 10-10 for Technical Support Instructions.SymbolsThere are special symbols that appear on the Precision Xceed Pro Monitor,

Pagina 82

12-3Questions? Refer to Page 10-10 for Technical Support Instructions.Bar Code Types Codabar: Codabar is a self-checking, numeric-only bar code. Codab

Pagina 83 - Bar Code Types

12-4Questions? Refer to Page 10-10 for Technical Support Instructions.RSS: RSS is an abbreviation for Reduced Symbology Set, which produces very small

Pagina 84 - Important Safety Instructions

12-5Questions? Refer to Page 10-10 for Technical Support Instructions.Standards Compliancee Precision Xceed Pro System has been evaluated against the

Pagina 85

1-5Questions? Refer to Page 10-10 for Technical Support Instructions.Calibration Verification ControlsUse RNA Medical brand Glucose and β-Ketone Calib