  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 71
10-8
Questions? Refer to Page 10-10 for Technical Support Instructions.
Symptom/Error Message Explanation How To Respond
The test result is below the measuring
limit of the system.
1. Repeat the test with a new test strip.
2. If the result persists, follow your facility's
procedure.
The test result is above the measuring
limit of the system.
1. Repeat the test with a new test strip.
2. If the result persists, follow your facility's
procedure.
The test strip or the test strip port is wet,
defective, contaminated or the
wrong test strip was inserted.
1. Remove the old test strip.
2. Press 1 to continue testing with a
New Strip.
Or,
Press 2 to Exit to the Test Menu.
ere may be a problem with the test strip.
One of several 4-digit error codes may
be displayed. ese codes provide Abbott
personnel additional information about the
problem. Errors include:
1. Press 1 to repeat the test with a New
Strip or 2 to Exit.
2. If the error occurs again, record the
4-digit error code and contact Abbott
Technical Support.
4327 - e strip was removed during
testing.
Repeat the test with a new test strip.
4330 - Blood glucose may be too high to
be read by the system or there may be a
problem with the test strip.
Repeat the test with a new test strip. If the
error occurs again, conrm the result by
performing a laboratory reference test.
20 mg/dL
During Any Test:
Chapter 10: Troubleshooting