8-1
Questions? Refer to Page 10-10 for Technical Support Instructions.
Monitor Configuration
The Precision Xceed Pro Monitor configuration settings are enabled via the data management system. These settings
can be viewed on the hand-held device, but cannot be changed on it. Settings that are configured through the data
management system are:
Upload interval
Length of time between data uploads.•
Whether you can continue to test (“allow test” or warn”) or not (“lockout”).•
Operator ID
Whether the Operator ID is required, optional or not used for all tests.•
Define the Operator ID prompt.•
The minimum and maximum number (between 1 and 30) of digits that are acceptable for the Operator ID. •
With manual entry of the Operator ID, determine whether a check digit is required (modulus 10 or 11).•
Decide acceptable formats for the Operator ID bar code. Refer to the Bar Code Specifications in Chapter 12 for •
available formats.
The monitor can truncate (ignore) the leading digits and/or the last digits of an Operator ID. The monitor can be set •
to select specific digits of an Operator ID. The number of digits is configured in the data management system
.
Patient Tests / Patient ID
Determine whether the Patient ID is required, optional or not used for patient tests.•
Define the Patient ID prompt.•
The minimum and maximum number (between 1 and 30) of digits that are acceptable for the Patient ID.•
Permission for manual entry of the Patient ID.•
Determine whether duplicate entry of a manually entered ID is required.•
With manual entry of the Patient ID, determine whether a check digit is required (modulus 10 or 11).•
Decide acceptable formats for the Patient ID bar code. Refer to the Bar Code Specifications in Chapter 12 for •
available formats.
The monitor can truncate (ignore) the leading digits and/or the last digits of a Patient ID. The monitor can be set to •
select specific digits of a Patient ID. The number of digits is configured in the data management system.
Decide whether confirmation of the Patient ID required.•
Determine how the ID will be confirmed.•
Decide whether comment code is required, optional or not used for patient tests that are in the acceptable range •
for the facility and the test strip.
Decide whether comment code is required, optional or not used for patient tests that are outside of the acceptable •
range for the facility and the test strip.
Control Tests
Set which levels of control solution are required, optional or not used by your facility. The monitor can be •
configured to run any combination of low, mid and high control solution tests.
Set whether control solution test results display as numeric or as Pass/Fail.•
Determine whether a comment code is required, optional or not used for control tests that are within the •
acceptable range for the facility and the test strip.
Decide whether a comment code required, optional or not used for control tests that are outside of the acceptable •
range for the facility and the test strip.
QC Interval
Determine how frequently QC tests need to be run. This can be set to a number of hours or up to three specified •
times during the day.
Decide whether the operator can continue to test (“allow testor “warn”) or not (“lockout”) when the QC interval •
is expired.
Review Setup
Chapter 8: Review Setup