Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
10
Important: If the test fails to start,
sufficient blood sample may not have
been applied to the test strip.
Discard the current test strip and
repeat with new strip.
10. Wait for the monitor to analyze the
sample and display the result
The monitor counts down the 20
seconds it takes to analyze the
sample, then displays the test result
11. Note the result and whether it falls
outside the action range
NOTE:
The Outside Action Range screen will
appear if the result is out of the action
range
- Repeat test to confirm result
- Notify M.D.
- Send specimen to laboratory
- Enter comment code
If the result is above or below the
action range, an up or down triangle
appears in front of the test result.
NOTE:
The Critical Range screen will appear if
the result is out of the critical range
- Repeat test to confirm result
- Notify M.D.
- Send specimen to laboratory
- Enter comment code
Test Procedure, Continued
What You See on the
Display
What You Do
Comments
NOTE:
The In-Range Action Range screen will
appear if the result is within this action
range
12. Remove the test strip from the