Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
7
Touch the blood drop to white target area at tip
of test strip while the “Apply Sample” prompt
appears in the display window.
Hold syringe in place until test begins.
The syringe can briefly touch the test strip
without affecting the test result.
Never add additional sample.
Discard strip and
use a new one if needed.
Do not force blood into the strip from the
syringe. This may cause blood to enter the port
protector and damage the meter.
Collecting Venous Blood Samples
With meter ready, obtain a sample:
Item Comment(s)
Collect the venous blood sample in a collection
tube containing sodium or lithium heparin,
ensuring that the test tube is completely filled.
Do not use collection tubes that contain fluoride
or oxalate.
If the blood is collected from an intravenous line,
clear the line before drawing the sample into a
syringe that contains sodium or lithium heparin.
Allow the sample to reach room temperature
before performing a test on the sample; use the
sample within 30 minutes of collecting it.
Invert the tube with the sample several times
immediately before taking a sample.
Use a disposable transfer pipette to obtain a
sample from the center of the collection tube.
Touch the blood drop to white target area at tip
of test strip while the “Apply Sample” prompt
appears in the display window.
Hold pipette in place until test begins.
The pipette can briefly touch the test strip
without affecting the test result.
Never add additional sample.
Discard strip and
use a new one if needed.