Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
13
6. Open the foil test strip packet at the
notch and tear up or down to
remove the test strip
7. With the contact bars facing up,
insert the test strip into the test strip
port until it stops.
The monitor will prompt the operator to
apply the desired control solution to
the test strip.
8. Verify that the QC solutions have
not been opened for more than 90
days. Gently invert the required
control solution bottle 3-4 times then
apply a drop to the target area on
the test strip.
The monitor will beep when the
sample is accepted. The sample
accepted screen will be displayed.
Recap the contract solution bottle
tightly.
The test starts automatically, as the
sample is accepted.
9. Wait for the monitor to analyze the
control solution and display the test
results.
The monitor counts down the seconds
it takes to analyze the sample (20
seconds), then displays the test result.
10. Control results appear as either
PASS or FAIL.
11. If the control result appears as FAIL,
scan or manually enter a comment
code via the keypad, then press
Enter.
12. Remove the test strip and discard.