Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
8
Test Procedure
Use the following procedure to perform patient tests.
Note: The word “Strip” on the display always refers to the test strip. If the monitor displays a message not
indicated in these procedures, please refer to section 10,
Troubleshooting, of the Abbott Precision Xceed Pro
manual
.
What You See on the
Display
What You Do
Comments
1. Press on/off to turn on monitor
The Abbott logo screen will appear
for a few seconds followed by the
product name, software version
screen and then the Test Mode
menu
2. Press 1 to select Patient Test
After the operator selects the type of
test, the display will prompt for an
Operator ID to be entered
3. Press Scan to scan the Operator ID
barcode, or manually enter the
Operator ID via the keypad, then
press
Enter
The display will next prompt for a
Patient ID to be entered
4. Press Scan to scan the Patient 9-10
digit CSN barcode, or manually
enter the Patient 9-10 digit CSN via
the keypad, then press Enter
After entering the Patient ID, the test
strip lot screen appears next.