Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
15
Docking Station
The Precision Xceed Pro™ System docking station provides a means for hands-free, automatic data transfer
(upload/download) between the Precision Monitor and a server in the Partners server farm running the QC
Manager Application software.
The lantronix box must be connected to a power source via the AC adapter supplied with the docking station.
Once the monitor is placed into the docking station and a connection is established, data will be automatically
transferred.
The circulating arrow, displayed on the meter’s LCD screen, indicates data transfer (upload/download) is in
process.
Do not remove meter until the arrow disappears and the menu is displayed.
Troubleshooting
A. Troubleshooting Out-of-Range Patient Test Results
This section describes conditions that can cause out-of-range patient test results. Refer to the Precision Xceed Pro
Test Strip package insert for test strip specification information.
NOTE: Do not use during intravenous infusion of high-dose ascorbic acid or during xylose absorption testing
Reasons Glucose Results May Be
Higher than Expected:
Hematocrit is lower than 20%
Serum or plasma samples instead of whole blood
Venous blood tested in arterial/capillary mode
Additional sample added to test strip during testing.
Reasons Glucose Results May Be
Lower than Expected:
Hematocrit is higher than 70%
Arterial or capillary blood tested in venous mode
Patient has hyperglycemic-hyperosmolar (with or without ketosis)
Patient has severe dehydration, hypotension, or shock
Water or alcohol remaining on the puncture site
Venous or arterial blood sample not tested within 30 minutes after collection
B. Troubleshooting Out-of-Range Control Solution Results
Repeat the test for that control solution to make sure that the operator meets the conditions in this checklist:
No air bubbles are in the control bottle’s nozzle.
Calibration performed using the bar-code for the test strip used.
Correct 5-digit lot number for the control solutions.
Solutions and test strips stored between 39°F and 86°F (4°C and 30°C).
Temperature and humidity conditions in the room where the tests are being performed are within the
acceptable operating range. Refer to the Precision Xceed Pro Test Strip Package insert.
Bottles have not been open for more than 90 days.
Only Precision Xceed Pro Test Strips used.
If test results are out of range despite meeting the above criteria, please repeat the test using a new box of control
solutions and/or test strips. If the results are still out of range, call:
Abbott Technical Support 1-877-529-7185
Nursing POCT Program Manager. 617-724-8763 (for PCS sites)
Pathology POCT Program Coordinators: 617-726-1462; 617-643-5392 (for PCS and other sites)