Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
19
This screen shows the level
of control test the monitor is
expected to run: Low or
High, and the level of the lot
entered.
Different level of control from the one
that the monitor expected
Press 1 to repeat the test.
Press 2 to run the level of test that
matches the lot number already
entered.
Troubleshooting When a Data Review is Selected
Symptom/
Error Message
Explanation
How to Respond
No stored test results for the type of
data requested.
Press 1 to return to the Data Review
Menu.
No stored test results match the
selected criteria.
Press 1 to return to the previous
screen.
Troubleshooting During Data Transfer
Symptom/
Error Message
Explanation
How to Respond
Error has occurred during data
transfer. This screen will only appear
for three seconds.
1. Remove the monitor from the
docking station.
2. Place the monitor back into the
docking station.
If the error appears again, contact the
POCT Coordinators.
Meter Replacement Procedure
In the event of a problem with the meter, the operator should:
1) Review the troubleshooting section of this procedure.
2) Contact the POCT Coordinators for assistance 6-1462, 3-5392 or 4-8763.
3) Email MGH POCT Coordinators
.
4) Contact the nursing supervisor for further assistance or additional issues on off-shift or weekends.
Competency Assessment
All operators must read the procedure manual and complete the “Training and Competency Record for the Precision
Xceed Pro Meter” during initial training. Competency is assessed at orientation and annually using at least two of the
following methods:
1. Successfully perform a low and high QC on the glucose meter.
2. Supervisor observes performance of routine work.
3. Each user’s quality control performance is monitored.
4. Written testing specific to the method.