Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
14
13a. After low control has been
successfully performed, press 1 –
Next Level and repeat steps 4-12
using high control.
13b. If high control fails, enter comment
code, press
Enter and repeat test
14a. When both levels PASS, the
analyzer is ready for patient testing
14b. If one or both levels continue(s) to
fail, take remedial action (see
appropriate section)
Documentation
Data is automatically retained by the meter and transferred to the QC Manager Computer during meter
docking. Contact POCT Program for detail.
Data Review
Patient History
The Precision Xceed Pro™ Monitor stores up to 4,000 patient tests that can be retrieved by the following
categories:
All Tests: This option displays all patient test results, starting with the most recent test results (reverse
chronological order).
Patient ID: This will display the test results for a particular patient, starting with the most recent test.
Operator ID: This will show test results performed by a particular operator, starting with the most recent test.
Action Range: This will display the test results that fall outside of the action range.
Comment Code: This shows the test results for all tests tagged with a specific comment code.
Date: This will display all the test results for a specific date, starting with the most recent test done on that
date.
Data Upload
Frequency: Periodic (as prompted by meter)
To begin data upload, place the monitor in the docking station. Once the connection is established,
communication between the monitor and the POCT Department PC will begin, automatically.
The monitor can send and receive data. When this occurs, the monitor will display the following message:
“Please wait for Data Upload”. Testing is preempted while the monitor is sending or receiving data.
The monitor must not be removed before data transmission is completed, otherwise data transfer will be
incomplete and operator access may be affected.