C. Discrepancies between Xceed Pro and Laboratory
Discrepancies noted between the Xceed Pro monitor and the Laboratory should be investigated. It is important to note
that the samples must be drawn at the same time or within a few minutes of each other to have a valid comparison as
blood glucose values can change quickly depending on the clinical condition of the patient. Accuracy of the glucose
results from a meter is considered acceptable if the glucose is within 15 mg/dL for results less than 75 mg/dL or +
20%
for glucose results greater than 75 mg/dL.
Nursing POCT Program Manager: 617-724-8763 (for PCS sites)
Pathology POCT Program Coordinators: 617-726-1462; 617-643-5392 (for PCS and other sites)
Email: MGH POCT Coordinators
D. Symptom/Error Messages
In this section, you will find information relating to error messages that appear on the display when errors are detected
by the Precision Xceed Pro™ Monitor.
For each corresponding message, an explanation is given and appropriate responses are described. In many
situations, it may be possible to proceed with some of the functions, at least temporarily, before attending to the
problem. (For example, when the batteries are too low to permit testing, it may be possible to review data for a short
time).
If any problem persists, record the error message displayed, including the three digit error code (if any), and call:
Abbott Technical Support 1-877-529-7185
Nursing POCT Program Manager. 617-724-8763 (for PCS sites)
Pathology POCT Program Coordinators: 617-726-1462; 617-643-5392 (for PCS and other sites)
Email: MGH POCT Coordinators
Troubleshooting When You Turn on the Meter, or During Use
Symptom/
Error Message
Explanation
How to Respond
The display is blank
The monitor has little or no power Place the monitor into the docking
station to recharge the battery or
change battery.
Battery power is getting low. Testing will
be available for a limited time.
Replace with 2 AA batteries.
The monitor may have a problem that
prevents if from operating properly. One
of several four-digit codes may be
displayed. These codes provide Abbott
additional information about the problem.
1. Turn off the monitor. Turn it on
and repeat the function.
2. If the problem persists, record
the three-digit error code and
contact the POCT program
manager for new meter.
An error has occurred during last data
transfer. This screen will appear when
you turn on the monitor.
Place the monitor into the docking
station to complete the upload.
Once the monitor has successfully
been docked, the warning will
disappear.
Or,
Press 1 to Continue testing.